a.
Slide the PEEK nut, stainless steel ring, and Super Flangeless
ferrule over the pump end of the steel tubing.
PEEK nut
b.
Insert the tubing in the pump, and tighten the fittings.
c.
At the sample selector valve, screw the natural-color PEEK
female-to-male adaptor into port 7.
d.
Slide the transparent, 1/16-inch flangeless nut and blue, 1/16-inch
flangeless ferrule over the tubing.
Rationale:
case of a blockage.
e.
Insert the tubing in the female-to-male adaptor in port 7, and
tighten the fittings.
5-118 Maintenance Procedures
Stainless steel ring
Flangeless nut
Flangeless ferrule
These components provide fail-safe pressure relief in
Super Flangeless ferrule
Stainless steel tubing
PEEK female-to-male
adaptor

Hide quick links: