To remove the existing capillary:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The probe and source can be hot. To avoid burn injuries, take
Warning:
great care while performing this procedure.
The ESI probe tip is sharp. To avoid puncture wounds, handle
Warning:
the probe with care.
1.
Remove the probe from the source (see
2.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 probe end cover retaining screws.
The probe and source components can be contaminated
Replacing the ESI probe sample capillary
page
3-6).
End cover retaining screws
5-67