Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual page 165

Hide thumbs Also See for Xevo G2 QTof:
Table of Contents

Advertisement

4.
Locate the PEEK coupler in the union, and carefully slide the
fused-silica capillary into the union until it butts against the pin-plug
capillary locator.
The pin-plug capillary locator correctly locates the capillary
Rationale:
in the union and ensures a minimum dead volume when fitting the
TaperTip.
5.
Tighten the PEEK coupler to hold the capillary in place.
6.
Unscrew and remove the pin-plug capillary locator.
Warning:
TaperTip that is sharp and fragile. To avoid puncture wounds,
handle the probe with care.
Caution:
TaperTip into the union and not the tapered tip.
7.
Align the PEEK coupler in the end of the union and slide the TaperTip
into the union until it butts against the fused-silica capillary.
Caution:
TaperTip and causes blockages.
8.
Finger-tighten the coupler to hold the TaperTip securely without
crushing.
Caution:
easily broken during insertion into the source.
9.
Mount the reference probe on the NanoLockSpray source.
10. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 reference probe retaining screws.
11. Using an F-130 Finger-tight Nut, connect the fused-silica capillary to
the grounded union in the instrument's IntelliStart Fluidics system.
12. Connect the reference probe's cable to the instrument's high voltage
connector.
Replacing the reference probe capillary (NanoLockSpray source)
The reference probe tip is an exposed, fused-silica
Ensure that you are inserting the square-cut end of the
Overtightening the coupler on the TaperTip crushes the
The fragile TaperTip at the end of the reference probe is
5-89