Parameter Alarm Level - GE Medical Systems Dash 2000 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2000:
Table of Contents

Advertisement

Parameter Alarm
Level
Revision C
Parameter alarms are assigned to one of the four patient status alarm
categories. The PARAMETER ALARM LEVEL option allows you to view
and reassign parameter alarms to other levels.
NOTE: The heart rate limit alarm level cannot be set to the MESSAGE
level in Neonatal ICU mode.
If you want to move a parameter alarm to another level, use the PARAM
ALARM LEVEL option, and follow the same procedure as described for
arrhythmia alarm levels. Any changes made with this menu option are
temporary and will revert to defaults upon discharge.
21-NOV-2000 16:27
II
PARAMETER ALARM LEVELS
>
RETURN
HR
WARNING
SPO2
ADVISORY
RR
ADVISORY
RR-APNEA
ADVISORY
SPO2-RATE
MESSAGE
NBP
MESSAGE
TP
MESSAGE
MAIN
MENU
PARAM.
PREV.
ALARM
MENU
LEVEL
PARAMETER ALARM LEVELS Popup Menu and
Dash 2000 Patient Monitor
2005873-001
DAK.BED 1
The parameters listed are
only those currently being
monitored.
048
Information Window
4-11

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dash 2000 and is the answer not in the manual?

Table of Contents

Save PDF