Patient Monitor Type

Admit Mode

Revision C
Following is an explanation of two Service Menu options which have been
mentioned in other places in the manual.
In many places throughout the manual the patient-monitor type is
mentioned. Your monitor is set up for monitoring in one of three modes
with the PATIENT-MONITOR TYPE option:
Adult-ICU (ADLT-ICU),
Neonatal-ICU (NEO-ICU), or
Operating Room (OP-ROOM).
You will know which mode your monitor is set up for by the absence or
presence of a label on the date and time line at the top of the display.
Adult-ICU has no label. The other labels are NEO-ICU and OP-ROOM.
The patient must be DISCHARGED to change the patient monitor type.
Remember, when you change the patient-monitor type any Monitor
Defaults you have set will be lost and factory defaults will again be in
effect.
Another option in the Service Menu which affects your monitor is the
ADMIT MODE. The admit mode determines the menu options you will use
when admitting a patient to the monitor. The admit modes are:
Standard,
Rover,
Combo, and
Rover Combo (ROV COM).