GE Medical Systems Dash 2000 Operator's Manual page 201

Patient monitor
Hide thumbs Also See for Dash 2000:
Table of Contents

Advertisement

Problem:
Erroneous NBP
measurement
Revision C
Solutions
1. Check for proper cuff size:
a. Too small a cuff can give an erroneously high value.
b. Too large a cuff can give an erroneously low value.
2. Check for residual air left in the cuff from a previous measurement.
3. Make sure cuff is not too tight or too loose.
4. Make sure cuff and heart are at same level, otherwise hydrostatic
pressure will offset the NBP value.
5. Minimize patient movement during measurement.
6. Watch for pulsus paradoxus.
7. Check for leak in cuff or tubing.
8. Patient may have a weak pulse.
9. Calibration may be necessary.
Dash 2000 Patient Monitor
2005873-001
10-17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dash 2000 and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF