General Cleaning

Exterior Surface

7-4
WARNING
Disconnect the monitor from the power line before cleaning or
disinfecting its surface.
The equipment should be cleaned on a regular basis. (Comply with the
policies of your institution's infection control unit and/or biomed department.)
The exterior surfaces of the equipment may be cleaned with a dampened, lint-
free cloth. Use one of the following approved solutions:
ammonia (diluted),
Cidex,
sodium hypochlorite bleach (diluted), or
mild soap (diluted).
To avoid damage to the equipment, follow these rules:
CAUTION
Failure to follow these rules may melt, distort, or dull the finish of
the case, blur lettering on the labels, or cause equipment failures.
Always dilute the solutions according to the manufacturer's
suggestions.
Always wipe off all the cleaning solution with a dry cloth after cleaning.
Never use a cleaning substance containing wax.
Never pour or spray water or any cleaning solution on the equipment or
permit fluids to run behind switches, into the connectors, or into any
ventilation openings in the equipment.
Never use these cleaning agents:
abrasive cleaners or solvents of any kind,
acetone
ketone,
alcohol based cleaning agents, or
Betadine.
Dash 2000 Patient Monitor
2005873-001