General Cleaning
Exterior Surface
7-4
WARNING
Disconnect the monitor from the power line before cleaning or
disinfecting its surface.
The equipment should be cleaned on a regular basis. (Comply with the
policies of your institution's infection control unit and/or biomed department.)
The exterior surfaces of the equipment may be cleaned with a dampened, lint-
free cloth. Use one of the following approved solutions:
•
ammonia (diluted),
•
Cidex,
•
sodium hypochlorite bleach (diluted), or
•
mild soap (diluted).
To avoid damage to the equipment, follow these rules:
CAUTION
Failure to follow these rules may melt, distort, or dull the finish of
the case, blur lettering on the labels, or cause equipment failures.
•
Always dilute the solutions according to the manufacturer's
suggestions.
•
Always wipe off all the cleaning solution with a dry cloth after cleaning.
•
Never use a cleaning substance containing wax.
•
Never pour or spray water or any cleaning solution on the equipment or
permit fluids to run behind switches, into the connectors, or into any
ventilation openings in the equipment.
•
Never use these cleaning agents:
abrasive cleaners or solvents of any kind,
acetone
ketone,
alcohol based cleaning agents, or
Betadine.
Dash 2000 Patient Monitor
2005873-001
Revision C