Pulse Oximetry (SpO2)

Non-invasive Blood Pressure (NBP)

14-28
Output specs (contd.):
Common mode
rejection:
Noise:
Accuracy:
Alarms:
Parameters monitored:
SpO2 range:
calibrated:
total:
PPR range:
Accuracy:
SpO2:
PPR:
Alarms:
Measurement technique:
Displayed parameters:
Measurement modes:
NBP pressure range:
Systolic pressure
range:
Adult:
Pediatric:
Neonatal:
Dash 2000 Patient Monitor
2005873-001
>60 dB at 60 Hz
<0.2 mmHg
p-p
±2% or ±1 mmHg (in the range of
–25 mmHg to 350 mmHg) whichever is
greater (exclusive of transducer)
User-selectable upper and lower limits
for systolic, diastolic, and mean
pressures
Arterial oxygen saturation (SpO2) and
peripheral pulse rate (PPR)
50 – 100%
0 – 100%
25 – 250 beats per minute (±3 beats per
minute)
Actual accuracy depends on probe.
Please reference manufacturer's
specifications.
±2% (70 – 100% SpO2) ±1 standard
deviation
±3% (50 – 69% SpO2) ±1 standard
deviation
±3 beats per minute
User-selectable upper and lower limits
for SpO2 and PPR
Oscillometric
Systolic, diastolic, and mean pressures,
pulse rate, time of last measurement
Manual, auto, and stat in Adult and OR
modes; manual and auto in Neonatal
mode
30 to 275 mmHg
30 to 235 mmHg
30 to 135 mmHg
Revision C