Setting Up The Measurement - Larson Davis SoundAdvisor 831C Reference Manual

Sound level meter
Hide thumbs Also See for SoundAdvisor 831C:
Table of Contents

Advertisement

In order to use the
TAKE NOTE
measurement counter, your
time needs to be greater than
00:01:00.
FIGURE 15-4
Timed Stop Mode
SoundAdvisor Model 831C
Interval Time
This is the time each measurement interval will be. The
minimum permitted Interval Time is one minute; if a zero
value is entered, it will default to one minute.
Interval Time Sync
This feature ensures that all measurement records, after
the first, will begin at a time of day equal to a multiple of
the measurement time selected. For example, if the
interval time is five minutes and the measurement
begins at 08:14:00 (hh:mm:ss format), the first measure-
ment will be cut short such that the subsequent mea-
surements will begin at 08:15, 08:20, 08:25, etc.
Daily Auto Store
As your measurement is being made, your measurement
can be stopped and stored automatically throughout the
day. For example, you can have an interval time of ten
minutes, and store 24/day, so every hour a file will be
created, and the measurement will be reset, and each
file will contain six measurement intervals.
Timed Stop Mode
This mode, when Measurement History is enabled, has a feature not
included in the other run modes: the ability to automatically measure
and store a user-defined number of records, then stop. Subsequent
runs, each manually initiated, will produce the same number of stored
measurements.
In this example the time is 00:05:00 and
the measurement counter is 5. This
means that the measurement will run
for 5 minutes, five times, each time
creating a new record.

Setting up the Measurement