5.3 Choosing a Calibrator

If using a 1/4 inch
TAKE NOTE
microphone an adaptor
ADP024 is required.
FIGURE 5-6
Larson Davis CAL200
SoundAdvisor Model 831C
Larson Davis recommends the following calibrator for the
SoundAdvisor:
Larson Davis Model CAL200: 94/114 dB @ 1 kHz
The following instances will require a correction:
Using Free Field Microphone
Check calibration data shipped with calibrator
The calibrator and instrument are at a temperature other than
room temperature (23° C) or not near sea level.
Check calibration data shipped with calibrator
The instrument is used in one environment and moved to
another (e.g. meter moved to or from 85 kPa static pressure
elevation)
Calibration check will need to be done in new environ-
ment.