11.3 Lock Modes

FIGURE 11-2
Lock Mode Menu
Unlocked
Lock with Auto Store
To learn more
LEARN MORE
about Big Digit, see "Big Digit
Sound Level" on page 4-3.
Lock with Manual Store
SoundAdvisor Model 831C
means that a code indicated as "0123" will be saved to the
meter as "123".
Step 4 Check or uncheck Allow Calibration when Locked.
Step 5 Close and Save.
There are three lock modes each with different degrees of security:
Refer to Table 11.1 on page 4 for constraints on the various lock modes.
The SoundAdvisor is completely unlocked and you can access all
features, data displays, and change any setting.
The data display will default to Big Digit and you cannot navigate to
any other views.
A measurement can run, stop, and store in this lock mode.
Measurement cannot be paused in this mode. When the measurement
is stopped, a dialog box will appear, "Measurement OK?" with Yes and
No buttons. Selecting Yes will save the measurement with an auto-
generated file name. Selecting No will cause a reset without saving.
If System Properties indicates an autostore prompt, you will be
prompted to change file name and save file. If you select No, the
measurement will stop and data will remain on the meter.
The data view will default to Big Digit and you cannot navigate to any
other views.
A measurement can run, stop, store, and reset in this lock mode.
If you've indicated an auto-store mode in your current setup file, it will
run and store as indicated in the setup.