14.2 Upgrading SoundAdvisor Firmware

The SoundAdvisor
TAKE NOTE
can be connected via USB or
via TCP/IP address, if the
meter is connected to the
internet. To learn more, see
"Communications & WiFi" on
page 13-1.
FIGURE 14-2
Upgrading Firmware/Options
SoundAdvisor Model 831C
G4 is used to implement the firmware upgrade and perform the
following steps:
Step 1 Turn the SoundAdvisor on.
Step 2 Connect the meter to the PC (or laptop) via the USB cable
CBL138, which is part of the PSA029 power supply.
Step 3 Launch G4.
Step 4 Navigate FileUpgrade Firmware or Options on G4.
Step 5 Select your meter, click Connect.
Step 6 After
communication
Upgrading Firmware/Options will appear.
Step 7 Check the radio button for Firmware.
Step 8 Select Choose Firmware File. The default folder will be
C:\Program Files (x86)\PCB Piezotronics\G4\Firmware
Step 9 Choose latest firmware from this folder or navigate to a
.fw831c firmware file, select and open file.
Step 10 Select Upload Firmware.
Upgrading SoundAdvisor Firmware
has
been
established,
the
14-2