11.5 Constraints
Table 11.1 Allowed Operations During Lock
Lock Mode
Lock with Auto Store
Lock with Manual Store
Fully Locked
11.6 Unlock
FIGURE 11-5
Unlock
SoundAdvisor Model 831C
Press the center softkey to navigate to the Calibrate display. For
information on how to calibrate, see "Calibration" on page 5-1.
Run
Pause
Yes
No
Yes
No
Yes
No
The SoundAdvisor can be accessed, locked/unlocked, settings can be
changed, and the measurement can run, pause, stop and store during a
lock when operated using G4 LD Utility.
To unlock the SoundAdvisor, follow these steps:
Step 1 Press the 3 Tools or 5 Enter key.
Step 2 To access keyboard, press 5 enter again.
Step 3 Enter your code, then press 5.
To unlock your SoundAdvisor via G4 LD Utility follow these steps:
Step 1 Connect your SoundAdvisor to G4.
See i.2.1 "Connecting SoundAdvisor to G4" on page i-3.
Constraints
Stop
Reset
Yes
Yes
Yes
Yes
No
No
Calibration
Yes (when stopped)
Yes (when stopped)
Yes (when stopped)
11-4
Need help?
Do you have a question about the SoundAdvisor 831C and is the answer not in the manual?