Pausing The Measurement - Larson Davis SoundAdvisor 831C Reference Manual

Sound level meter
Hide thumbs Also See for SoundAdvisor 831C:
Table of Contents

Advertisement

7.3.1 Pausing the Measurement

A measurement
TAKE NOTE
may be paused and resumed
multiple times.
Erase Previous 5-10 Seconds of Measurement
Pressing any other
TAKE NOTE
key besides the back-erase
option will end the back-erase
state.
FIGURE 7-4
Back Erase
The Back Erase
feature is located
where Menu
typically is.
Undo appears as an
option after a ten
second erase.
After the first erase,
you can erase an
additional 5 seconds
(not an additional 10).
SoundAdvisor Model 831C
At any time the measurement of overall data can be temporarily
paused. During a pause, the run clock will not stop and instantaneous
data will continue to be displayed on the Live display.
After a pause, press the 9 RUN/PAUSE key to resume the
measurement. The overall data will not be affected by any acoustic
events occurring during the time period that the SoundAdvisor was
paused.
The SoundAdvisor has a back-erase feature that allows for a quick
erase of the last 5 to 10 seconds of measured data. This feature is used
to erase sounds that are recorded, but not wanted in the overall
measurement.
Enable Back-Erase Feature
The back erase is visible on the Overall page, when the measurement is
paused, and only after the measurement has run more than 5 seconds
since the last run or resume.
The back erase feature is unavailable when any of the following have
been enabled in the Active setup:
Measurement History
Event History
Spectral Ln
Event Sound Recording
Session Log
The session log will have an icon to indicate data had been erased.
Performing the Measurement
To close Back Erase,
start a run again or Stop
the measurement.