25. The sample chamber should be loose in the assembly at this point. If not, back out the sleeve
until the sample chamber is loose. Use the end of a 0.156 Allen wrench in the slot of the sleeve
to adjust to the point of eliminating the longitudinal play of the sample chamber.
26. Tighten the sleeve in 24 1/4-turn increments (6 turns total) to preload the seals.
27. Lightly lubricate the inside bore of the cylinder.
28. Align the fittings on the cylinder with the sample chamber tubes, or with air lines if servicing.
29. Slide the cylinder onto the piston and the body.
30. Install two cone point set screws into the cylinder and tighten into the groove on the body.
31. Assemble the vaporizer assembly around the vaporizer chamber.
32. Insert the four screws that retain the two parts of the vaporizer assembly and tighten them.
33. Install the insulation around the vaporizer assembly.
34. Install the insulation cover over the vaporizer assembly and insulation.
35. Open the isothermal oven door.
36. Orient the flange so that the marks made at LSV removal are lined up. This should have the
sample chamber vertical with the air connections at the top of the valve and the connection of
the vaporizer chamber at the top.
37. Install the four flange screws and tighten them.
38. Reinstall tubing to the LSV, being careful to connect each line to the proper connection on the
LSV.
39. Close the isothermal oven door.

Removing the LSV from the Analyzer

1. Stop the analysis at the end of a cycle.
2. Turn off power, carrier, sample, and air to the analyzer.
3. Open the isothermal oven door.
4. Label all tubing connections to the Liquid Sample Valve.
5. Disconnect all tubing from the Liquid Sample Valve.
6. Mark the analyzer and flange to ensure the flange is reinstalled in the correct orientation.
7. Remove the four screws holding the flange to the side of the analyzer.
8. Remove the LSV from the analyzer.
9. Remove the insulation cover and insulation from around the vaporizer assembly.
10. Remove the four screws that retain the two parts of the vaporizer assembly.
8-8
2000-OSM, F1