Control Parameters Screen - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

3. On the Control Parameters screen (see Figure 6-20), press the F3 (Alarm Mask) soft key.
4. On the Alarm Mask, Main screen (see Figure 6-21), cursor down to the desired alarm name and
verify the selection in the right-hand column.
5. To change the alarm mask setting, toggle the left or right cursor to obtain the desired selection
(Alarm, Masked, Indicate, or Warning).
2000-OSM, F1
*****
Control Parameters
Analyzer Name:
[ABB Analytical GCC
Unknown Peak Alarm:
Missing Comp Alarm:
Update Trends on Alarm:
Auto Update of RF:
Max RF Variation:
Unknown Peak RF:
Unknown Peak Units:
Exit
Escape
and
Update
F1
F2
Figure 6-20. CONTROL PARAMETERS SCREEN
*****
Alarm Mask, Main
Active Sensors:
(Expand)
Component High:
Component Low:
Hardware:
Power Fail:
Unused:
Idle:
Flame Out:
Clock:
Expand
Exit
Mask
F1
F2
Figure 6-21. ALARM MASK MAIN SCREEN
]
No
No
No
Yes
No
0.000000
0.000000
%
More Below
Alarm
Mask
Out
F3
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
(Expand)
Alarm
More Below
F3
*****
Yes
Yes
Yes
ppm
Common
Digital
Mask
F4
*****
F4
6-41