Preparing A Method Table - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

between the various components is established (at installation), only the K-factor is changed
during calibration.

Preparing a Method Table

This subsection explains how to prepare a method table by constructing a sample method table.
Figure 4-12 shows a chromatogram of the analysis to be controlled. This analysis is 90 seconds long
and has three components to measure: A, B, and C. These components elute at the following times
(after the start of the analysis): component A, 30 seconds; component B, 50 seconds; and
component C, 70 seconds.
In order to construct a method table for this analysis, you need to know when to turn valves on and
off, how the components are defined, and how to measure the components. For this analysis we
shall use slope detection, which is described in detail in "PEAK DETECTION." Figure 4-13 illustrates
the data needed to construct a method table for this analysis.
TIME-RELATED DATA
Time
Function
0001
Valve On
0005
Valve Off
0006
Stream Step
0007
Auto Zero
0012
Slope Detect
0020
Valve On
0021
Valve Off
0030
Component
0040
Valve On
0041
Valve Off
0050
Component
0060
Valve On
0061
Valve Off
0070
Component
0089
Slope Detect
To construct this method table perform the following steps:
1. On the Background screen, press the F1 (Exit to Commands) soft key.
4-15
A
0
TIME (in seconds)
Figure 4-12. CHROMATOGRAM OF ANALYSIS
Value
One
One
N/A
N/A
On Now
Two
Two
One
Three
Three
Two
Two
Two
Three
Off Now
Figure 4-13. DATA FOR METHOD TABLE
C
B
Method Table No.: 1
Analysis Cycle Time: 90 sec
Calibration type: external std
Analysis Normalization: no
Primary Cal Stream: 10
Secondary Cal Stream: not used
Benchmark Stream: not used
Multiplier Exponent: 0
K Factor: 1
Component Units: PPM
Key Component: #2
Retention Time Deviation: ±5 sec
Response Factor: 1
required
Calibration Concentration: 100
for each
Benchmark Concentration: 100
component
High Concentration Alarm: 100
Low Concentration Alarm: 0
Calculate By: area
In Primary Cal Stream: yes
90
OTHER DATA
2000-OSM, F1