Display Screen; Checking Calibration - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Display Screen

This analyzer operates from a series of menus and screens, using the display on the GCC. The
Background screen is the default screen, from which you can select the other screens. Since all
screens are divided into the same basic elements, you should understand this screen structure before
operating the analyzer (see Figure 3-4).
**********
ABB Analytical GCC
803V029-1R (Std GCC)
**ALARM**
**ALARM**
**ALARM**
Analysis Time :00114
Current Stream :01
Purging Stream :01
Exit
to
Commands
Figure 3-4. BACKGROUND SCREEN WITH ELEMENTS IDENTIFIED
When the Background screen is not displayed, you can return to the Background screen by
repeatedly pressing the F1 soft key until the Background screen appears.
Before you start into full analytical operation, perform a calibration (if appropriate) and a single
analysis on each stream. These precautions enable you to verify the analyzer and the process.
When you stop analytical operation, select "At End of Analysis" unless an
emergency occurs. Selecting "Now" stops an analysis immediately, which can
inadvertently damage the analyzer if it stops while an analysis is in progress.
Stopping during an analysis can trap sample in a column or permit temperature or
pressure ramps to continue without adequate control.

Checking Calibration

If you have a calibration run specified for your analyzer, request a calibration run and compare the
results with those in the Data Package. To do this perform a single calibration run:
2000-OSM, F1
B A C K G R O U N D
20-JUN-2002 THU 08:16:01
Manual
Control
F1
F2
**********
Remote 0.1
(C) 2001
Peaks Found : 00
Key Component:
none
entered
Cur Report,
Basic I/O &
Alarms
Options
F3
CAUTION
Message Line
Soft Keys
F4
Screen Title
Data Display
Soft Key
Labels
3-3