Background Screen - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

1. Verify that the Background screen is displayed (see Figure 2-12). If it is not displayed, press the
F1 soft key several times until the Background screen appears.
2. Press the F1 (Exit to Commands) soft key.
3. On the Commands screen cursor down to STORAGE & CONFIG.
4. Press the F2 (Save or Restore Tables) soft key.
5. On the Table Storage screen cursor down to RESTORE TABLES FROM E2PROM.
6. Press the F2 (Restore Tables) soft key.
7. When "Are You Sure? Y/N" appears on the screen, press the "Y" key.
8. Press the F1 soft key as necessary to return to the Background screen.
This restores to main memory the tables that were most recently saved to E
If ABB Lewisburg provided Applications Engineering, verify all method tables by comparing them to
those in the Data Package. This is done most easily by printing the tables (if a printer is connected to
the analyzer). To print the method tables:
1. Display the Background screen.
2. Press the F1 (Exit to Commands) soft key.
3. On the Commands screen cursor down to PRINTER.
2-20
**********
B A C K G R O U N D
ABB Analytical GCC
803V029-1R (Std GCC)
**ALARM**
**ALARM**
20-JUN-2002 THU 08:16:01
**ALARM**
Analysis Time :00114
Cycle Time
:00000
Current Stream
:01
Purging Stream
:01
Method
:01
Exit
Manual
to
Control
Commands
F1
F2
Figure 2-12. BACKGROUND SCREEN
**********
Remote 0.1
(C)
2001
Peaks Found:00
State:Idle
Key Component:
none
entered
Cur Report,
Basic I/O &
Alarms
Options
F3
F4
2
PROM.
2000-OSM, F1

Advertisement

Table of Contents
loading

Table of Contents