Location Of Zone Fittings - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

3. Eliminate any back pressure to the zone by loosening the feedthrough fitting associated
with the zone to be calibrated (see Figure 8-23).
4. On the Controller, verify the Background screen is displayed and then press the F2
(Manual Control) soft key.
5. On the Manual Control Mode screen press the F4 (Press. & Temp. Control) soft key.
2000-OSM, F2
Figure 8-23. LOCATION OF ZONE FITTINGS
8-51