MATCHING COLUMN FLOWS

When you are matching column flows, refer to the Data Package for flow rates and valve
configurations. Although methods of matching column flows vary with valve configurations, the
following procedures apply to most configurations. Verify flows using a flow meter connected to the
point being checked (e.g., cell vent).

Manual Operation of Valves

Operate the valves manually as follows:
1. On the Background screen press the F2 (Manual Control) soft key.
2. On the Manual Control Mode screen (see Figure 2-5), the first line ("Valves:") indicates which
valves are ON. The cursor will be on the second line ("Enter Valve: 0"), which enables you to
turn valves on and off.
3. To turn Valve One ON, verify that the number "1" does not appear on the first line, then press the
"1" key on the keyboard. The number "1" appears momentarily on the second line and then
appears on the first line.
4. To turn Valve One OFF, verify that the number "1" appears on the first line ("Valves:"). Move the
cursor to the second line and press the number "1" key. The number "1" disappears from the first
line of the screen.
5. Turn Valve Two and other valves on and off manually in the same manner.
2-10
******
Manual Control Mode
Valves:- - - - - - - -
Enter Valve:
[0]
RECORDER:
Offset
:
0 %
Attn
:
5
Off Chroma Bar Trend
PURGE SELECT:
0
Start
Crest
Time
Time
Last:
52
55
38
43
Exit
Detector
to
and
Background
Graphics
F1
F2
Figure 2-5. MANUAL CONTROL MODE SCREEN
******
Time:00000
Cycle Time:00000
Method Tbl:
Stream:
Purging:
Peaks:
00
State: Idle
End
Peak
Time
Area
Type
73
18032
Val-Bas
51
15322
Bas-Val
Manually
Press. &
Zero
Temp.
Baseline
Control
F3
F4
000
01
01
2000-OSM, F1