VAC Conditioning PCB

There can be either one or two VAC Conditioning PCBs, depending on the application. Each board
has two jumpers to be set, JP4 and JP5 (see Figure 8-30). These jumpers are set according to the
voltage supplied to the board, as follows (JP4 is the hot side and JP5 is the neutral side):
Applied Voltage
32 VAC
110 VAC
220 VAC
Figure 8-30. VAC CONDITIONING BOARD JUMPER LOCATIONS

Front Panel PCB

There is one adjustment and three jumpers on this board (see Figure 8-31). The adjustment is
potentiometer R56, which is used to set the display contrast for optimum viewing.
Figure 8-31. FRONT PANEL PCB ADJUSTMENT AND JUMPER LOCATIONS
The jumpers relate to the LOCOM option. JP5 has its shunt is between pins 1 and 2 for normal
operation, between pins 2 and 3 for LOCOM installations. JP6 and JP7 have the shunts in place for
normal operation and the shunts removed for LOCOM installations.
8-58
JP4 Shunt
JP5 Shunt
1-2
1-2
3-4
3-4
5-6
5-6
JP4
JP5
JP7
JP6
JP5
R56
2000-OSM, F1