Common Do Mask Main Screen - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

4. On the Common DO Mask, Main screen (see Figure 6-24), cursor down to the desired alarm
name and verify the selection in the right-hand column.
5. To change the alarm common output, toggle the left or right cursor to obtain the desired selection
(Common DO or Masked).
6. If the selected alarm has "(Expanded)" after the alarm name, the alarm has subordinate
indicators. To view these subordinates, cursor to the desired alarm line and press the F2
(Expand Mask) soft key.
7. On the expanded alarm screen, repeat steps 4 and 5 to change a specific alarm's output.
8. Press the F2 (Exit and Update) soft key to save the new alarm data, or press the F1 (Escape) soft
key to retain the original data.
9. Press the F1 soft key as necessary to return to the Background screen.
6-44
*****
Common DO Mask, Main
Active Sensors:
(Expand)
Component High:
Component Low:
Hardware:
Power Fail:
Idle:
Flame Out:
Clock:
Method Function: (Expand)
Expand
Exit
Mask
F1
F2
Figure 6-24. COMMON DO MASK MAIN SCREEN
Common DO
Common DO
Common DO
Common DO
Common DO
Common DO
Common DO
(Expand)
Common DO
Common DO
More Below
F3
*****
F4
2000-OSM, F1