Peak Inversion; Chromatogram Oversized Or Off Scale - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Cause: The gating in the method needs to be modified to ensure gating of the missing peak in the
report. This can be done by modifying the gate ON/OFF time in the method table. See "PEAK
DETECTION" in Section 4 for detailed instructions.

Peak Inversion

Symptom: The peaks on the strip chart recorder are reversed or negative (see Figure
6-13).
Cause: Negative peaks on the strip chart recorder could be due to reversed recorder wiring at the
analyzer or at the recorder. Some recorders are equipped with polarity reversal. Ensure correct
polarity setting on recorder. Other possible causes include reversal of sense and reference tubing
connections or filament wiring if a TCD is used, or a grounded jet or missing jet voltage if a FID is
used.
1. Check for reversed sense and reference filament wires.
2. Check for reversed sense and reference tubing connections to the detector.
3. Make necessary corrections.
4. Perform Test Procedure 28 (FID Polarizing Voltage Test).

Chromatogram Oversized or Off Scale

Symptom: The chromatogram or peak areas are excessively large. The condition may be severe
enough that all peaks are off scale on the highest attenuation (see Figure 6-14). Chromatogram may
go off scale at approximately the time of distillation on and not return to scale until time of distillation
off.
Cause: The attenuation could be set too low, making the chromatogram or peaks appear large (loss
of polarizing voltage on FID). Another possible cause is the wrong carrier gas is being used (e.g.
nitrogen used instead of helium).
6-10
NORMAL
Figure 6-13. PEAK INVERSION
NORMAL
Figure 6-14. CHROMATOGRAM OFF SCALE
Baseline offset
adjusted up scale
OFFSET
PEAK OFFSCALE
2000-OSM, F1

Advertisement

Table of Contents
loading

Table of Contents