Gas Control (Analog) - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Temperature Control
Method:
Ambient Oven
Temperature:
Setpoint Resolution:
Temperature Stability:
Steady Ambient:
Ambient Range:

Gas Control (Analog)

Control Method:
Filtration:
Inlet Pressure:
Minimum:
Maximum:
Supply pressure must be less than or equal to 50 psig (345 kPa, 3450 mbar) higher
than the maximum output required (e.g., supply to a 100 psig (689 kPa, 6.89 bar)
regulator should not exceed 150 psig (1034 kPa, 10.34 bar). It is the customer's
responsibility to set relief devices used to protect the analyzer and prevent
hazardous releases should be set at a maximum of 200 psig (1380 kPa, 13.80 bar).
Range:
Gauges:
Temp Coefficient:
Regulation:
Accuracy:
Repeatability:
Allowable Gasses:
Quality:
Flow Adjustment:
Tube Fittings:
2000-OSM, F1
Closed loop PID
+ 30° to 180° C (Settings and display in °C only)
1° C
±0.1° C
±1.0° C
Mechanical regulators
2 microns at inlet, provided
Setpoint + 20 psig (138 kPa, 1.38 bar)
150 psig (1034 kPa, 10.34 bar)
NOTE
2-15 psig
0-15 psig
(0-102 kPa)
(0-1.02 bar)
0.3 psi/77°F (2.1 kPa/25°C)
0.01 psi (0.7 kPa) outlet / 1.0 psi (6.9 kPa) inlet
N/A
N/A
H
, He, N
, Air, Ar
2
2
GC Grade
Oven mounted needle valves or flow controllers with external
adjustment
316 SS Gyrolok (standard)
316 SS Swagelok (optional)
1/16, 1/8, 1/4 inch connections
5-30 psig
0-30 psig
(0-200 kPa)
(0-2.00 bar)
20-100 psig
0-100 psig
(0-700 kPa)
0-7.00 bar)
1-5