Gas Control (Digital) - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Gas Control (Digital)

Control Method:
Number of Zones:
Filtration:
Inlet Pressure:
Minimum:
Maximum:
It is the customer's responsibility to set relief devices used to protect the analyzer
and prevent hazardous releases at a maximum of 200 psig (1380 kPa, 13.80 bar).
Range:
Gauges:
Temp Coefficient:
Regulation:
Accuracy:
0 to 50 psig:
50 to 100 psig:
Repeatability:
Allowable Gasses:
Quality:
Flow Adjustment:
Tube Fittings:
1-6
Closed loop PID; temperature stabilized
1 to 5
2 microns at inlet, provided
Setpoint + 10 psig (69 kPa, 0.7 bar)
100 psig (689 kPa, 6.89 bar)
NOTE
0 to 100 psig (0 to 689 kPa, 0 to 6.89 bar), bubble tight, non-
venting
Electronic readout: 0.01 psig (0.07 kPa, 0.7 mbar) resolution
Setpoint resolution: 0.01 psig (0.07 kPa, 0.7 mbar)
None
See below
1.7%
2.7%
±0.1 psig (±0.7 kPa, ±7 mbar)
H
, He, N
, Air, Ar
2
2
GC Grade
Oven mounted needle valves or flow controllers with external
adjustment
316 SS Gyrolok (standard)
316 SS Swagelok (optional)
1/16, 1/8, 1/4 inch connections
2000-OSM, F1