4. On the Manual Temperature Control Mode screen (see Figure 4-38), select the desired field
using the Up and Down cursor keys. Enter two digits for valve numbers (e.g. 04), as some
installations have more than nine valves.
ISO-OVEN
PROG. OVEN
T/C CELL
LSV
Valve # : [0]
Zone # : 2
Set Point: 220.00
Ramp Rate: 30.0000
VALVES : - - - - - - - -
Figure 4-38. MANUAL TEMPERATURE CONTROL MODE SCREEN
5. After you enter changes in the Set Point or Ramp Rate fields, press the F2 (Accept Control Point)
soft key to save the new data.
Data changes made on the screen go to the DTC Backplane PCB, which compares the new settings
with the application-specific temperature limits. If the new settings fall within the limits, it accepts
them and applies them to the analysis.
If entries in the Set Point or Ramp Rate fields fall outside the set limits, the * TEMP BD ALARM *
flashes on the screen's message line. Temperature Zone specific alarms also cause the zone name
to flash. To identify the type of alarm, press Escape to return to the Pres & Temp Control Screen,
then press the F4 (Alarms) soft key to display the Alarm Screen. After you note the type of alarm
triggered and consult the Data Package for the set limits, clear the alarm, return to the Manual
Temperature Control Screen, and change the entries to values within the set limits. If the DTC
Backplane PCB accepts the new settings, there will be no * TEMP BD ALARM * display on the
screen after you press the F2 (Accept Control Point) soft key.
Zero All Zones and Span All Zones settings are made in the factory and are not intended to be
changed.

5 Zone Names and Limits Table

This screen allows you to enter low and high alarm limits and change the temperature zone names.
Always check the original application-specific temperature limits listed in the analyzer's Data Package
to validate any changes. To access the 5 Zone Names and Limits Table perform the following steps:
1. On the Background screen, press the F2 (Manual Control) soft key.
2. On Manual Control Mode screen, press the F4 (Press. & Temp. Control) soft key.
4-41
****
Manual Temperature Control Mode
Zone
Config
ISO
PROG/C
SWITCH
ISO
Accept
Escape
Control
Point
F1
F2
Actual
SP
79.9C
80.0C
220.1C 220.0C 30.0
199.8C 200.0C 10.0
199.9C 200.0C 10.0
Time : 0125
State : Pk Back
T-Rating
T2
Zero
Span
All
All
Zones
Zones
F3
****
Rate
0.0
Ver
290C
F4
2000-OSM, F1