8. On the Request Analysis screen (see Figure 2-13), select the method number for a calibration
run.
9. Place the cursor on "Cal" and then press the F2 (Accept The Request) soft key.
10. Press the F1 (Escape) soft key to return to the Commands screen.
11. If the analyzer is running, go to step 14. If the analyzer is stopped, go to step 12.
12. On the Commands screen, press the F4 (Start/Stop Analysis) soft key.
13. On the Start/Stop Analysis screen, press the F2 (Start Analyzer) soft key. The controller will
perform the calibration as soon as it completes all other requests.
14. Press the F1 soft key as necessary to return to the Background screen.
15. Watch the screen to determine when the calibration stream (noted in step 4) has purged and
starts the calibration.
16. Press the F3 (Current Report, Basic I/O, and Options) soft key.
17. On the Report, Basic I/O and Options screen, cursor to CURRENT REPORT, if necessary, and
press the F2 (View Current Report) soft key. The desired report will display at the end of the
calibration.
18. When the report is complete, press the F1 soft key as necessary to return to the Background
screen. The analyzer will continue to run until the operator stops it.
19. To stop the analyzer, press the F1 (Exit to Commands) soft key.
2-22
******
Request Analysis
Analysis Requests:
or calibration enter the method#,
otherwise enter the stream#:[1
None
Stream
PENDING:
00
Accept
Escape
The
Request
F1
F2
Figure 2-13. REQUEST ANALYSIS SCREEN
For a benchmark
]
Bench
Cal
00
00
Clear
All
Requests
F3
F4
******
2000-OSM, F1