8. On the Temperature Control screen, move the cursor as necessary to select the Zone Name from
among the names in the list, and to change Set Point and Ramp Rate values of the desired
entries.
When you exit the Temperature Control screen, the value in the Entry line
highlighted when you exit becomes the Value entry in the Method Table. Unless
you want to change the Value entry, exit on the same Entry line number that you
entered the Temperature Control screen.
9. Verify the changed values and place the cursor on the desired Entry number. Then press the F1
(Exit) soft key.

Temperature Check Table

This screen allows you to select zone names and change alarm temperature values and alarm
methods, all related to the time they occur in the analysis method. To access this screen perform the
following steps:
1. On the Background screen, press the F1 (Exit to Commands) soft key.
2. On the Commands screen, cursor to TABLE EDIT and press the F2 (Methods) soft key.
3. On the Methods Table #?? screen, enter the number of the Method Table to be changed and
press the F2 (Edit the Table) soft key.
4. On the Cal Define & Cycle Time screen, press the F1 (Continue) soft key.
5. On the Method Table screen (see Figure 4-41), cursor to the appropriate line in the Time field, or
press the Insert Line soft key to insert a new line (just below the cursor) and enter the time into
the cycle when the function is to be performed.
6. Cursor to the Function field and scroll to Temp Check. In Figure 4-41, with Temp Check in the
Function field the Value column lists One as the function to be performed from the Temperature
Check Table.
2000-OSM, F1
CAUTION
4-46