Remote Number Identification With Sw1 - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

5. Set the switch positions to the desired remote number, using Figure 2-15 as a guide.
Figure 2-15. REMOTE NUMBER IDENTIFICATION WITH SW1
6. Insert the Chroma I/O PCB into slot 2 of the Card Cage and make sure the PCB is securely in
place.
7. Close and lock the GCC front panel.
8. Turn power to the analyzer ON at the circuit breaker.
The local control options are manual control, view only, and screen saver, using SW1 positions 6, 7,
and 8 (SW1 positions 6 and 7 interact).
Setting SW1-7 to OFF locks the screen in a "view only" mode; you cannot edit any tables or
functions. Setting SW1-7 to ON disables the "view only" function.
SW1-6 provides manual control but does not let you update tables. Setting SW1-6 to OFF
enables manual control, but only when SW1-7 is in the ON position. When SW1-7 is set of OFF
this overrides the setting of SW1-6.
2000-OSM, F1
Switch No.
5 4 3 2 1
1 1 1 1 1
#1
#2
1 1 1 1 0
1 1 1 0 1
#3
#4
1 1 1 0 0
#5
1 1 0 1 1
#6
1 1 0 1 0
#7
1 1 0 0 1
1 1 0 0 0
#8
1 0 1 1 1
#9
1 0 1 1 0
#10
1 0 1 0 1
#11
1 0 1 0 0
#12
1 0 0 1 1
#13
1 0 0 1 0
#14
1 0 0 0 1
#15
#16
1 0 0 0 0
Switch No.
5 4 3 2 1
0 1 1 1 1
0 1 1 1 0
0 1 1 0 1
0 1 1 0 0
0 1 0 1 1
0 1 0 1 0
0 1 0 0 1
0 1 0 0 0
0 0 1 1 1
0 0 1 1 0
0 0 1 0 1
0 0 1 0 0
0 0 0 1 1
0 0 0 1 0
0 0 0 0 1
0 0 0 0 0
Remote No.
#17
#18
#19
#20
#21
#22
#23
#24
#25
#26
#27
#28
#29
#30
#31
#32
2-25

Advertisement

Table of Contents
loading

Table of Contents