8. On the Pressure Control screen, move the cursor as necessary to select the Zone Name from
among the names in the list, and to change Set Point and Ramp Rate values of the desired
entries.
When you exit the Pressure Control screen, the Entry line highlighted when you
exit becomes the Value entry in the Method Table. Unless you want to change the
Value entry, be sure to exit on the same Entry line number on which you entered
the Pressure Control screen.
9. Verify the changed values and place the cursor on the desired Entry number. Then press F1
(Exit).

Pressure Check Table

This screen allows you to select zone names and change alarm pressure values and alarm methods,
all related to the time they occur in the method. The Pressure Check Table is reached from the
Method Table as follows:
1. On the Background screen press F1 (Exit to Commands).
2. On the Commands screen cursor to TABLE EDIT and press F2 (Methods).
3. On the Methods Table #?? screen, enter the number of the Method Table you want to change
and press F2 (Edit the Table).
4. On the Cal Define & Cycle Time screen, press F1 (Continue).
5. On the Method Table #01 screen (see Figure 4-49), cursor to the appropriate line in the Time
field, or press the Insert Line soft key to insert a new line (just below the cursor) and enter the
time into the cycle when the function is to be performed.
6. Cursor to the Function field and scroll to Pres Check. In Figure 4-49, with Pres Check in the
Function field the Value column lists One as the function to be performed from the Pressure
Check Table.
2000-OSM, F1
CAUTION
4-58