Replacing A Regulator; Replacing A Gauge; Installing The Carrier Gas Panel - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

11. Carefully remove the Carrier Gas Panel.

Replacing a Regulator

1. Remove the Carrier Gas Panel from the analyzer.
2. Disconnect the lines from the back of the regulator.
3. Remove the regulator adjusting knob.
4. Remove the nut securing the regulator to the panel.
5. Remove the regulator from the panel.
6. Remove the adjusting knob from the new regulator.
7. Insert the new regulator through the hole in the panel.
8. Fasten the nut securing the regulator to the panel.
9. Install the adjusting knob on the regulator.
10. Reinstall the lines on the back of the regulator.
11. Install the Carrier Gas Panel in the analyzer.

Replacing a Gauge

1. Remove the Carrier Gas Panel from the analyzer.
2. Disconnect the line from the back of the gauge.
3. Loosen the screws on the plate that secures the gauge to the panel and then remove the plate.
4. Remove the gauge from the panel.
5. Insert the new gauge through the hole in the panel.
6. Install the plate that secures the gauge to the panel.
7. Tighten the screws holding the gauge and plate to the panel.
8. Reinstall the line on the back of the gauge.
9. Install the Carrier Gas Panel in the analyzer.

Installing the Carrier Gas Panel

1. Insert the Carrier Gas Panel into the Control Housing access area of the analyzer controller.
2. In the Control Housing access area, install the five socket head screws into the manifold (see
Figure 8-19), making sure the sealing gasket is in place.
3. In the Control Housing access area, install the plugs into the unused ports of the manifold (see
Figure 8-19).
2000-OSM, F1
8-41

Advertisement

Table of Contents
loading

Table of Contents