Plumbing Installation; Gas Requirements; Calibration Sample - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Equipment:

Plumbing Installation

All plumbing connections enter at the right side of the cabinet. Be careful to avoid damaging fittings
and to ensure tight connections. Refer to the Data Package for specific plumbing information.
For brass and stainless steel pipe fittings, wrap the threads with Teflon tape or a suitable thread
sealant. Make all connections and then tighten them appropriately. Leak check each connection as
described in the "Leak Check" paragraph.

Gas Requirements

The following information represents general requirements. Due to differences in applications, refer
to the Data Package supplied with your analyzer for application-specific requirements.
Instrument Air: instrument grade air.
Burner Air: ISA Hydrocarbon Free Grade with hydrocarbon content less than 1 ppm and dew
point less than -40° C.
Carrier Gases: varies with the application; the lower the levels of components being measured,
the higher the necessary carrier gas purity.
TCD: for percent level measurements, gases of 99.99% purity; for ppm measurement ranges,
99.995% purity.
FID: for measurements down to 5 ppm, gases of 99.995% purity; lower ranges require higher
levels of purity (refer to the Data Package for specifics).
FPD: varies with the application; refer to the Data Package for specifics.

Calibration Sample

A primary factor in determining the accuracy of the analyzer's calibration is the quality and precision
of the calibration sample. Regardless of the sample composition or number of complete analysis
components, generally only measured components are in the calibration sample.
2-2
3/16-inch flat blade screwdriver
1/8-inch flat blade screwdriver
9/64-inch ball driver
7/16-inch open end wrenches (2)
1/2-inch open end wrenches (2)
9/16-inch open end wrenches (2)
5/8-inch open end wrenches (2)
6-inch adjustable wrench
12-inch adjustable wrench
3/8-inch drive socket set with ratchet handle and leverage bar
5/16-inch (8 mm) hex key wrench (supplied with Analyzer)
5/64-inch (2 mm) hex key wrench (supplied with Analyzer)
Leak testing solution (e.g., Snoop®)
Flow Meter
Stop Watch (if not using a digital flow meter)
Chart Recorder
Flashlight
2000-OSM, F1