9. On the Pressure and Temperature Control screen (Figure 7-5), cursor up or down until TEMP
CONTROL is highlighted. Press the F2 (Manual Temp Control) soft key to access the Manual
Temperature Control Mode screen (see Figure 7-7).
Figure 7-7. MANUAL TEMPERATURE CONTROL MODE SCREEN
10. Compare the set point values of temperature zones 1 through 5 with those of the factory table
listing that was provided with the chromatograph. If the values agree, there is no need to modify
the table. Exit this screen by pressing the F1 (Escape) soft key and proceed to step 12.
11. If the values do not agree, the values have been changed since the analyzer was shipped from
the factory. The correct values should be entered.
12. Return to the background screen by repeatedly pressing the F1 soft key until the Background
screen is displayed.
13. From the Background screen press the F1 (Exit to Commands) soft key.
7-22
****
Manual Temperature Control Mode
Zone
Config
ISO-OVEN
ISO
PROG. OVEN
PROG/C 220.1C 220.0C 30.0
T/C CELL
SWITCH 199.8C 200.0C 10.0
LSV
ISO
Valve # : [0]
Time : 0125
Zone # : 2
State : Pk Back
Set Point: 220.00
Ramp Rate: 30.0000
VALVES : - - - - - - T2 290C
Accept
Escape
Control
Point
F1
F2
Actual
SP
79.9C
80.0C 0.0
199.9C 200.0C 10.0
T-Rating
Ver
Zero
All
Zones
F3
****
Rate
Span
All
Zones
F4
2000-OSM, F1