2. On the Commands screen, cursor down to TABLE EDIT and then press the F2 (Methods) soft
key.
3. The Method Table #?? screen will appear (see Figure 4-14). Use the number keys on the GCC
front panel to enter the number "1" for this method (the method number given in Figure 4-13
data). When the correct number shows on the screen, press the F2 (Edit the Table) soft key.
4. The Cal Define & Cycle Time screen will appear (see Figure 4-15). Set the analysis cycle time
for 90 seconds using the number keys.
2000-OSM, F1
*****
Method Table #??
Enter the table number: [1]
Edit
Escape
The
Table
F1
F2
Figure 4-14. METHOD TABLE #?? SCREEN
******
Cal Define & Cycle Time
Analysis Cycle Time: [300] Sec
Calibration Type
Ext Std
Normalized Indirect Std
Analysis Normalization
Primary Cal Stream:
Secondary Cal Stream:
Benchmark Stream:
Multiplier Exponent:
K Factor:
1.0000000
Continue
F1
F2
Figure 4-15. CAL DEFINE & CYCLE TIME SCREEN
F3
No
Yes
10
0
0
0
F3
*****
F4
******
F4
4-16