Get Time Vs. Boiling Point Table Command; Goto Statement; Go To Subroutine Statement; Hardware Alarm Function - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Comments:
temperature! is the temperature to convert to a time.
time_vs_bp_table% is the Time vs. Boiling Point Table (1-8) to use for the
conversion.
Example:
0010 TIME! = TEMP_TO_TIME(212.0,1)
Returns the time that 212.0 degrees occurred.

GET TIME VS. BOILING POINT TABLE Command

Description:
Returns one of the eight Time vs. Boiling Point Tables (Sim Dis analyzers only).
Syntax:
GET_TVSBP_TBL(array!,table%)
Comments:
array! is a previously dimensioned array that is the destination of the Time
vs. Boiling Point Table.
table% is the table number (1-8).
Example:
0010 N = LEN_TVSBP_TBL(1)
0020 DIM A!(N)
0030 GET_TVSBP_TBL(A!,1)
Gets the size of the Time vs. Boiling Point Table, dimensions an array to
hold it, copies the table into array A!.

GOTO Statement

Description:
Branches to a numbered statement.
Syntax:
GOTO line
Comments:
line is the statement number to branch to.
Example:
0010 GOTO 30
0020 END
0030 30 LPRINT "BYE"
Skips statement 20 by branching to statement 30, prints BYE, then ends
executions.

GO TO SUBROUTINE Statement

Description:
Branches to subroutine.
Syntax:
GOSUB line
RETURN
Comments:
line is the statement number for the subroutine.
Example:
0010 GOSUB 100
0020 END
0030 100 RETURN
Branches to subroutine at statement 100. Subroutine returns execution at
statement 20 which ends execution.

HARDWARE ALARM Function

Description:
Returns the state of GCC Hardware Alarms.
Syntax:
HARDWARE_ALARM
Example:
0010 IF HARDWARE_ALARM GOTO 100

HIGH CONCENTRATION ALARM Function

Description:
Returns the state of GCC High Concentration Alarms.
Syntax:
HIGH_CONC_ALARM
Example:
0010 IF HIGH_CONC_ALARM GOTO 100
10-16
2000-OSM, F1