Trend And Detector; Basic Common! Size; Final Steps - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

Trend and Detector

1. Cursor to Other on the second Digital Configuration screen and press the F2 (Trend & Detector)
soft key.
2. On the Trend And Detector Config screen (see Figure 5-12), verify each trend setting and
detector configuration.
*****
Chroma Board 1
Det #
1
2
Figure 5-12. TREND AND DETECTOR CONFIGURATION SCREEN
3. If all entries are correct, press the F1 (Escape) soft key to return to the second Digital
Configuration screen. If there are any errors on this screen, correct the errors and then
press the F2 (Exit and Update) soft key to save the corrected data and return to the
second Digital Configuration screen.

BASIC COMMON! Size

1. Cursor to Other on the second Digital Configuration screen and press the F3 (BASIC COMMON!
Size) soft key.
2. On the BASIC COMMON! Size screen, verify the COMMON! size.
3. If it is correct, press the F1 (Escape) soft key to return to the second Digital Configuration screen.
If it is not correct, enter the correct size and then press the F2 (Exit and Update) soft key to save
the corrected data and return to the second Digital Configuration screen.

Final Steps

After all configuration changes are complete, you must save the tables to E2PROM and reset the
unit.
1. On the second Digital Configuration screen, press the F1 (Exit) soft key.
2000-OSM, F1
Trend And Detector Config
Trend 1
00 11 21 12 22
Trend 2
00 11 21 12 22
DETECTOR CONFIGURATION
None FID FPD T/C M1 M2 M3
None FID FPD T/C M1 M2 M3
Exit &
Escape
Update
F1
F2
*****
More Below
F3
F4
5-13