Checking Backflush Flow

1. Connect the flow-measuring device to the Backflush Vent and verify that Valves One and Two are
OFF (see Figure 2-8).
2. Adjust the Backflush Vent restrictor to the flow rate stated in the Data Package.
CARRIER