Power

The Neutral and Ground (Earth) connections to the PGC2000 must be at earth (0
Volts) potential. Failure to maintain earth (0 volts) potential at these connection
points constitutes a serious safety hazard.
Voltage and Frequency:
Power Consumption:

Instrument Air

Supply Connection:
Supply Pressure:
Quality:
Flow Rates, all purge types:
Start-up Purge:
Steady State Purge :

Analytical Detectors

Standard Detectors:
Configuration:
Third-party detectors:

Isothermal Analytical Oven

Oven Liner:
Internal Dimensions:
Number of Valves:
Columns:
Heat:
1-4
WARNING
100 VAC (+15, -6 VAC), 50/60 Hz ± 10%
120 VAC ± 10%, 50/60 Hz ± 10%
230 VAC ± 10%, 50/60 Hz ± 10%
1,200 VA Startup, 900 VA Steady-State Operation
(Typical, varies with installed options)
3/8 inch tube, minimum
60 psig (414 kPa, 4.14 bar) minimum
100 psig (689 kPa, 6.89 bar) maximum
Instrument grade: Clean, Oil Free and -34° C, (-30° F) dewpoint
3
7.6-8.5 ft
/min (214-242 L/min) at 60°F (20°C)
3
4.5-5.2 ft
/min (127-147 L/min) at 60°F (20° C)
Thermal Conductivity, Flame Ionization, Flame Photometric,
Inter-column Thermal Conductivity
Up to two simultaneous detectors, each with capability for
sequential inter-column TC detector (one FPD per analyzer)
Consult Factory for availability
Stainless Steel
15.3 in. W x 20.4 in. H x 9.0 in. D
(390 mm W x 520 mm H x 230 mm D)
Standard provisions for five gas samples or column switching
valves in oven; standard provisions for two external liquid sample
valves (consult factory for special requirements)
1/16, 1/8, 3/16 inch, Packed Stainless or Fused Silica Capillary
Forced Air
2000-OSM, F1