15. Perform FID cell maintenance for cleaning jet and flame out thermocouple (see "FID
REPAIR/Cleaning the FID Jet" in Section 8).
16. Replace the plastic shield, cover, and thumb nuts on the Detector Amplifier.
17. Close and tighten the Controller front door.

TP 30: CHECKING TEMPERATURE TABLE VALUES

Verify that the software and hardware temperature parameter values are correct and have not been
inadvertently changed. To perform this operation, it is necessary to access the temperature
configuration table and the manual temperature control mode.
1. Press the F1 soft key until the Background screen is displayed (see Figure 7-3).
2. On the Background screen press the F2 (Manual Control) soft key.
2000-OSM, F1
**********
B A C K G R O U N D
ABB Analytical GCC
803V029-1R (Std GCC)
**ALARM**
**ALARM** 18-JUN-2002 TUE 08:16:01
**ALARM**
Analysis Time
:00114 Peaks Found:00
Cycle Time
:00000 State:Idle
Current Stream
:01
Purging Stream
:01
Method
:01
Exit
Manual
To
Control
Commands
F1
F2
Figure 7-3. BACKGROUND SCREEN
Remote 0.1
(C) 2001
Key Component:
none
entered
Cur Report,
Basic I/O &
Options
F3
**********
Alarms
F4
7-19