Performing The Breathing Maneuver For The Mvv Test - NDD EasyOne Air Operator's Manual

Hide thumbs Also See for EasyOne Air:
Table of Contents

Advertisement

Breathing maneuvers for all available tests
7.4.

Performing the breathing maneuver for the MVV test

Operator's Manual V03 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
With the maximum voluntary ventilation test, the patient fully inhales and
fully exhales repeatedly.
Required material
A nose clip
Prerequisite
For this test, you must enter the environment temperature within 1°C or
1.8°F accuracy.
1 Tell the patient to breathe at rest.
2 Tell the patient to fully exhale.
3 Tell the patient to take the EasyOne FlowTube into the mouth and to seal
the lips around the EasyOne FlowTube.
 The patient must not block the opening with the tongue or teeth or bite
down excessively on the EasyOne FlowTube.
4 Tell the patient to fully inhale and fully exhale repeatedly for an
uninterrupted period of at least 12 seconds.
5 Tell the patient to take the EasyOne FlowTube out of the mouth and to
breathe normally again.
 Performing spirometry  4 1
9000007/03
61
60/105

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents