3.3.

Setting up the physical components

Operator's Manual V03 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
WARNING
ELECTRIC SHOCK
Patients and technicians may be exposed to dangerous voltage.
Connect only medical-grade power supply released by the manufacturer.
To ensure electric safety, connect only equipment, such as printers and net-
works, that complies with the IEC 60950-1 standard for the electrical safety of
IT equipment.
CAUTION
ELECTRIC SHOCK, PATIENT HEALTH HAZARD, AND FALSE DIAGNOSIS
Damage to casing or broken components may expose the patient or technician
to dangerous voltage. Small parts from damaged disposables can be ingested
or inhaled by the patient. Measurement errors can cause false diagnosis.
Check the device and especially the battery for visible damage before on-site
installation and before usage.
1 Unlock and open the battery compartment by turning the lock screw
clockwise.
2 Insert the battery into the battery compartment of EasyOne Air.