Electromagnetic Compatibility (EMC)

A.1.1

General

A.1.1.1

Environment

A.1.1.2

EMC conformance

A.1.2

Safety information

A.1.3

Compliant cables and accessories

Operator's Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
See:
 Intended use of the EasyOne Air spirometer, 2
EasyOne Air is EMC-tested in conformity with the requirements of
rd
IEC 60601-1-2:2007 3
th
IEC 60601-1-2:2014 4
EasyOne Air is a medical device that requires special safety precautions and
must be installed and placed in operation in accordance with the information
given in tables below.
See:
 Safety information regarding electromagnetic compatibility, 14
WARNING
INCREASED EMISSIONS OR DECREASED IMMUNITY
The use of accessories, transducers and cables other than those specified
may result in increased emission or decreased immunity performance of the
equipment or system.
The table below lists cables, transducers, and other applicable accessories
with which the manufacturer claims EMC compliance.
Accessories that do not affect EMC compliance are not included in the
following list.
USB cable micro to type A, shielded, 1.8 m
USB cable micro to type B, shielded, 1.8 m
ndd power supply 5.0 V DC
 Accessories with which the manufacturer claims EMC compliance
 List of box contents, 17
edition (see the following tables) and
edition (according clause 7 and 8.9, tables 4 to 9).