Calibration check

12
Calibration check
12.1

About calibration check with EasyOne Air

Operator's Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
CAUTION
FALSE DIAGNOSIS
A number of reasons can lead to malfunction of EasyOne Air, and malfunction
of EasyOne Air can lead to false results and false diagnosis.
Perform calibration checks periodically, according to ATS/ERS.
Calibration check, 118
Checking for correct operation of EasyOne Air, 122
MALFUNCTION
Calibration of EasyOne Air is not possible. You can merely check EasyOne Air
for correct calibration.
Do not attempt to repair EasyOne Air yourself. A third-party service must not at-
tempt to repair EasyOne Air either.
Do not open or remove the casing.
If a calibration check fails, contact only your EasyOne dealer or the
ndd Servicing Department for repairs.
Contact information, 4
There are three types of calibration checks: single cal check, multi-flow cal
check, and linearity cal check.
Single cal check
Single cal check is not standard-compliant. The manufacturer recommends
multi-flow cal check and linearity cal check.
With single cal check, you simulate one inspiration and one expiration in a
single trial. The measured volume must be within ±3.5% accuracy. You can
perform multiple trials.
EasyOne Air displays exclusively the best trial. But all trials are stored. The
printout shows the best three trials.