Oem Interface Panel; Installing The Host Module; Routing The Host Module Cable - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Removing and Installing the Host Module
7. Disconnect the cable from the connector J36.
8. Remove four screws from the bottom of the host module.
9. Disconnect all of the cables from the upper side of the host module.
10. Disconnect all of the cables from the OEM subpanel located on the left side
11. Disconnect all the cables from the OEM interface panel shown in Figure 4-
12. Loosen the four cap screws located at the rear of the host module.
13. Remove the host module from the G50 or the G60 S.

9.3 Installing the Host Module

1. Verify that the cables are securely positioned in both slits located at the
Page 42 of 52
Figure 4-38 OEM Interface Panel
of the host module.
38.
sides of the opening shown in Figure 4-39.
Figure 4-39 Routing the Host Module Cable
7482255 Rev 01
October 2002
Siemens

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sonoline g60