Siemens SONOLINE G50 Service Manual page 148

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Installing or Updating the DIMAQ-IP Software
1. Back up all exam and patient data to a CD before continuing with the
!
2. On the G60 S, save the feature key for the Stress Echo option by perform-
3. Open the upper-rear panel of the system to expose the Microsoft Windows
4. Record the 20 characters of the Microsoft Windows 2000 Product Identifi-
5. Close the upper-rear panel.
6. Press the F4 key and select the Service menu item.
7. Access the Service software according to the instructions in "Service Soft-
8. Select the DIMAQ-IP Maintenance button located on the Main Menu
9. Insert the CD containing the version 1.2 DIMAQ-IP software, in the CD
10. Move the mouse pointer to the bottom of the screen to display the Windows
11. Select Shutdown → Restart from the menu.
12. The DIMAQ-IP will reboot from the software CD; the display of the BIOS
13. The blue Radisys logo will be displayed.
14. The software installation requires approximately 5 minutes: The R:\>
15. After the R:\> prompt appears, remove the software CD.
16. Operate the power switch to shut down the system: After two minutes, if the
17. After the system has powered down, wait 15 seconds and re-energize the
Page 8 of 42
DIMAQ-IP software update.
Caution: Failure of the DIMAQ-IP software installation or update will not nor-
mally destroy presets or service data. As a precaution, this data can be backed
up according to the procedures provided in "Backup/Restore" (page 8-36) and
at step 3. (backing up presets) on page 8-2.
ing the procedure "Uninstalling the Product Feature Keys" (page 8-14).
Note: The Stress Echo feature key must be reinstalled after the DIMAQ-IP soft-
ware is updated. See "Installing the Product Feature Keys" (page 8-10).
label attached to the DIMAQ-IP.
cation code; this license code is required to enable the new software.
ware Operation" (page 8-17).
screen (Figure 8-10) to display the Windows 2000 screen running on the
DIMAQ-IP module.
drive.
toolbar, and select the Start button.
Note: The DIMAQ-IP will boot from the CD drive if Windows is restarted with a
bootable CD installed in the CD drive.
and Windows boot screens will be followed by the software installation win-
dow.
prompt appears when the installation is completed.
system has not powered down, operate the circuit breaker located at the
rear of the system to de-energize it.
system.
7482255 Rev 01
October 2002
Siemens

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sonoline g60

Table of Contents