Chapter 7 - Installing the Version 1.2 Software Update on the G60 S
6. Insert the MO disk that contains the appropriate backup into the MO drive:
7. Select the Go button located at the bottom of the screen to begin the
!
8. Eject the MO disk from the MO drive.
9. Select the Home button and the Exit button to exit the service software.
2.0 Verifying the Operation of the Software
After the G60 S version 1.2 software update is completed, perform the
following steps to verify that the software is operational.
1. Confirm that no warning message is displayed on the monitor.
2. Press the F4 key to display the Preset screen.
3. Use the trackball to select the Service menu item.
4. Press SET.
5. At the service password entry screen, verify that the correct software ver-
6. At the System Information screen, verify that the correct system Serial No.
The installation of the version 1.2 software update is complete.
Siemens
See "Backing Up Service Data" (page 7-2).
restore process.
Caution: Selecting the Cancel button during an incomplete service data restore
procedure will leave the system inoperable: Do no select the Cancel button
when restoring service data.
sion is displayed.
Note: Serial numbers appear in two formats. Serial numbers assigned to Euro-
pean systems carry the prefix of GAE; other serial numbers carry the prefix of
GAA. Both types of serial numbers contain a four digit number in addition to the
prefix.
is displayed: The number displayed on the System Information screen
should be identical to the system Serial Number recorded on the label
attached to the rear of the system.
7482255 Rev 01
October 2002
Page 7 of 8