Testing The Dimaq-Ip; Installing Dimaq-Ip And Host Module Cables; Removing The Dimaq-Ip Module - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Installing the G50 DIMAQ-IP
42. Attach the DIMAQ-IP option label at the designated option-label location.

1.4 Testing the DIMAQ-IP

Use the following steps to test the DIMAQ-IP module.
1. Connect the G50 power cord to the outlet, and energize the G50 system.
2. Verify that the G50 initializes without displaying error messages.
3. Press the Review key.
4. Verify that the DIMAQ-IP Digital Lab screen is displayed.
5. Verify that the DICOM Screen menu item is listed on the Change Screen
6. Operate the trackball and verify that the pointer moves across the screen.
7. Use the Digital Lab features to process several studies.

1.5 Removing the DIMAQ-IP Module

See "Removing and Installing the DIMAQ-IP Module" (page 4-24) for remove
and replace procedures appropriate for both the G50 and the G60 S DIMAQ-IP
module.
Page 10 of 16
*2
*4
*5: VGA cable
Figure 5-10 Installing DIMAQ-IP and Host Module Cables
menu located at the lower-left side of the screen: This indicates that the
DIMAQ-IP software is version 1.2 or greater.
*1
*3
*5
Host Module
7482255 Rev 01
October 2002
Siemens