Removing The Probe Interface Panel; Installing The A36 A3Rd-2 Pcb - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Chapter 5 - Installing Equipment Options
14. Discard the blank panel and the two screws that were removed from it.
15. Orient the A36 A3RD-2 board shown in Figure 5-13 to the A36 PSEL PCB.
16. Align the connectors on the A36 A3RD-2 PCB to the PSEL PCB.
17. Press the connectors of the A36 A3RD-2 PCB into the matching connec-
18. Install the six M3 X 8 screws, provided in the kit, through the A36 A3RD-2
19. As indicated in Figure 5-14, reinstall the probe interface panel on the front
Siemens
Remove two screws
from the reverse
to remove the
blank panel
"Do not open"
label
Figure 5-12 Removing the Probe Interface Panel
tors (P20, P21, P22, and P23) of the PSEL PCB.
PCB and into the PSEL PCB as shown in Figure 5-13.
P23
P22
Figure 5-13 Installing the A36 A3RD-2 PCB
of the G50, using the twelve M4 X 8 screws that were removed in step 11.
Pull the loop to open
A36 PSEL PCB
P21
P20
P21
P23
P22
P20
7482255 Rev 01
October 2002
Remove these
CK4 X 8
screws (12) from
the front side.
Probe Interface Shield
"Do not open"
label
Connect P20, P21, P22, and P23
when ins talling A36 A3RD-2 PCB
onto A36 PSEL PCB.
36 A3RD-2 PCB
Fasten
CK3 x 8 screw (6)
supplied with this kit
Probe Interfa ce Shield
Fasten
CK4 x 8 screw (2)
supplied with this kit
after installing Probe
Interface Shield
Page 13 of 16