Installing Or Updating The System Software - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Installing or Updating the System Software

1.0 Installing or Updating the System Software
Use the following general instructions to install and test the system software on
the G50 and the G60 S.
The original release of the G60 S contained version 1.0 system software and
version 1.022 DIMAQ-IP software. The Speed Info 011/2002/US describes the
update of the version 1.022 DIMAQ-IP software to version 1.023.
The original release of the G50 contains version 1.2 system software and
version 1.2 DIMAQ-IP software.
1.1 Installing the Software
The following steps describe the installation of the system software version 1.2
on the G50 and the G60 S.
!
1. Energize the G50 or the G60 S system.
2. Back up the existing service data according to the instructions provided in
3. Use the following procedure to back up all presets, QuickSets, OB tables,
Page 2 of 42
Note: See "Installing the Version 1.2 Software Update on the G60 S" (page 7-1)
for specific instructions covering the installation of the version 1.2 system soft-
ware update on the G60 S system.
Caution: Currently, the version number of the DIMAQ-IP software installed on a
system cannot be verified at the system; this bug will be remedied soon.
On existing G60 S systems, the version 1.0 system software will operate accept-
ably with both the version 1.023 DIMAQ-IP software and the version 1.2
DIMAQ-IP software; to avoid problems, update the version 1.022 or 1.023
DIMAQ-IP software to version 1.2, before updating the system software to ver-
sion 1.2. See "Installing or Updating the DIMAQ-IP Software" (page 8-7) for this
procedure.
Note: On G60 S systems using the version 1.0 software, a formatted MO diskette
is required for the preset and service data backup operation because the MO
drive was incapable of formatting a diskette. On systems using version 1.2 the
formatting of MO diskettes is supported.
"Backup/Restore" (page 8-36).
Note: Refer to the "Documentation and Storage" chapter of the System Refer-
ence manual for the preset backup procedure.
and formulas:
a. Press the Review button to view the Study Screen.
b. Select the MO Disk radio button.
c. Select the Preset tab.
d. Select the Store Quicksets button.
e. Check the All Presets checkbox and select the Ok button.
7482255 Rev 01